EEC ESA-150A Electrical Safety Compliance Analyzer (0 - 4500W)
Manufacturer: EEC Model: ESA-150A Guarantee: 12 Month - Request a Quotation
- Contact
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222)730.39.68
HCM city: (028) 38.119.636
INPUT
• Voltage (AC): 115/230V ± 15% Auto Range
• Frequency: 50/60 Hz ± 5%
AC WITHSTAND VOLTAGE
• Output Rating: 5kV/100mA
• Output Voltage Range: 0-5.00kV
• Voltage Resolution: 0.01 kV
• Voltage Accuracy: (1% of setting + 0.5% of Range)
• Current Measurement Range (Total): 0.000-100.00mA
• Current Resolution (Total): 0.001/0.01mA
• Current Accuracy (Total): ±(2% of reading + 2 counts)
DC WITHSTAND VOLTAGE
• Output Rating: 6kV/20mA
• Output Voltage Range: 0-6.00kV
• Voltage Resolution: 0.01kV
• Voltage Accuracy: ±(1.5% of setting + 5V)
• Current Measurement Range: 0.0μA-20.00mA
• Current Resolution: 0.1μA/0.001mA/0.01mA
• Current Accuracy: ±(2% of reading + 2 counts)
INSULATION RESISTANCE
• Output Rating: 1kV/50GΩ
• Output Voltage Range: 30-1000V
• Voltage Resolution: 1 V
• Voltage Accuracy: ±(1.5% of setting + 2 counts)
• Resistance Measurement Range: 0.050MΩ-50GΩ
• Resistance Resolution: 0.001/0.01/0.1/1MΩ
• Resistance Accuracy:
0.050-999.9MΩ under 30-499V: ±(7% of reading+2 counts)
0.050-999.9MΩ under 500-1kV: ±(2% of reading+2 counts)
1G-9.999GΩ under 500-1kV: ±(5% of reading+2 counts)
10G-50GΩ under 500-1kV: ±(15% of reading+2 counts)
GROUND BOND
• Output Rating: 40 A/600mΩ/8V
• Output Current: 1.00-40.00A
• Current Resolution: 0.01A
• Current Accuracy: ±(2% of setting + 2 counts)
• Output voltage: 3.00-8.00V
• Voltage Resolution: 0.01V
• Voltage Accuracy: ±(2% of setting + 3 counts)
CONTINUITY
• Output Rating (DC): 0.1A for 0-10.00Ω, 0.01A for 10.1-100.0Ω, 0.001A for 101-1kΩ, 0.0001A for 1.001-10kΩ, 0.1A is Max.
• Resistance Offset: 0.00-10.00Ω
• Resistance Offset Resolution: 0.01Ω
• Resistance Offset Accuracy: ±(1% of reading + 3 counts)
• Resistance Measurement Range: 0.00-10kΩ
• Resistance Resolution: 0.01/0.1/1Ω
• Resistance Accuracy:
0.00-10.00Ω/10.1-100.0Ω: ±(1 % of reading + 3 counts)
101-1000Ω/1001-10000Ω: ±(1 % of reading + 10 counts)
TOUCH CURRENT
• Leakage Current Range (RMS): 0.0μA-10.00mA
• Leakage Current Resolution (RMS):
0.0-999.9μA: 0.1μA
1000-8399μA: 1μA
8.40-10.00mA: 0.01mA
• Leakage Current Accuracy (RMS) (AC + DC):
DC: ±(2% of reading + 3 counts)2
15Hz < f <100kHz: ±(2% of reading + 3 counts)2
100kHz < f < 1MHz: ±(5% of reading) (> 10.0μA)
RUN TEST
Power Measurement Range: 0 - 4500W
Power Accuracy: ± (5% of reading + 3 counts)
Power Factor: 0.000 - 1.000
Power Factor Accuracy: ± (8% of reading + 2 counts)
Voltage Measurement Range(AC): 0.0 - 277.0V , 1ø
Voltage Accuracy: ± (1.5% of reading + 2 counts)
Current Measurement Range(AC): 0.00 - 16.00A
Current Accuracy: ± (2% of reading + 2 counts)
Leakage Current Measurement Range: 0.00 - 10.00 mA
Leakage Current Accuracy: ± (2% of reading + 2 counts)
MD (L-G) Resistor: 2kΩ ± 1%
GENERAL
• Interface: Standard USB & RS232, Optional Ethernet, GPIB
• Display: 800 x 480 resolution digital TFT LCD/Contrast 9 Levels 1-9
• Dimension (W x H x D), mm: 430 × 133 × 500
• Weight: 36 Kg
- Quality Engagement
- Easy change and return
- Delivery Avaliable
- Favorable payment