FLUKE 1652C Multifunction Installation Tester

AC Voltage Measurement: 500 V/ ± (0,8% + 3 digit)
Continuity Testing: 20 Ω, 200 Ω, 2000 Ω/± (1,5%+3 digit)
Insulation Resistance Measurement:
+ Test Voltage: 250 V, 500V, 1000V
+ Test Current: 20 Ω, 200 Ω, 2000 Ω/± (1,5%+3 digit)
+ Accuracy: ± (1.5% + 3 digits)
Loop Impedance Measurement: 100 - 500 VAC (50/60 Hz)/ ± (3% + 6 digits) in no Trip mode,  ± (2% + 4 digits) in Hi Current mode
PFC, PSC Test: 0 - 25 kA/  Determined by accuracy of loop resistance and mains voltage measurements.
RCD Testing 
Tripping Time Test (ΔT)
Tripping Current (ramp): 50% - 110% of RCDs rated current/±5%
Phase Sequence Indication 
Size (L x W x H): 10 x 25 x 12.5 cm
Weight 1.3 kg
Batteries: Type AA: 6

Details

Manual

Specifications

AC Voltage Measurement

Range:

 500 V

Resolution:

 0.1 V

Accuracy (50 - 60 Hz)

 ± (0.8% + 3 digits)

Input Impedance

 3.3 MΩ

Overload Protection

 660 Vrms

Continuity Testing

Range (autoranging):

 20 Ω, 200 Ω, 2000 Ω

Resolution:

 0.01 Ω, 0.1 Ω, 1 Ω

Test Current:

 > 200 mA

Open Circuit Voltage:

 > 4 V

Accuracy:

 ± (1.5%+3 digits)

Insulation Resistance Measurement

Test Voltage:

 250 - 500 - 1000 V

 

 

Test Voltage:

 250 V

Test Current:

 1 mA @ 250 kΩ

Range/Resolution:

 20 MΩ/0.01 MΩ

Range/Resolution:

 200 MΩ/0.1 MΩ

Accuracy:

 ± (1.5% + 3 digits)

 

 

Test Voltage:

 500 V

Test Current:

 1 mA @ 500 kΩ

Range/Resolution:

 20 MΩ/0.01 MΩ

Range/Resolution:

 200 MΩ/0.1 MΩ

Accuracy:

 ± (1.5% + 3 digits)

Range/Resolution:

 500 MΩ/1 MΩ

Accuracy:

 10%

 

 

Test Voltage:

 1000 V

Test Current:

 1 mA @ 1 MΩ

Range/Resolution:

 20 MΩ/0.01 MΩ

Range/Resolution:

 200 MΩ/0.1 MΩ

Accuracy:

 ± (1.5% + 3 digits)

Range/Resolution:

 500 MΩ/1 MΩ

Accuracy:

 10%

 

 

Auto Discharge:

 Discharge time constant, 0.5 second for
C = 1 µF or less

Live Circuit Detection:

 Inhibits test if terminal voltage > 30 V prior to initiation of test

Maximum Capacitive Load:

 Operable with the 5 µF load

Loop Impedance Measurement (High Current Mode and Non-Trip Mode)

Range:

 100 - 500 VAC (50/60 Hz)

Input connection:

 Soft Key selection

Loop impedance:

 Phase to earth

Line impedance:

 Phase to neutral

Limit on consecutive tests:

 Automatic thermal shutdown after 50 consecutive tests at 10 second intervals (typical)

 

 

Max. Test Current. High Current and Non-Trip Modes:

 

Maximum Test Current @ 400 V

 20 A sinusoidal for 10 ms

Maximum Test Current @ 230 V

 12 A sinusoidal for 10 ms

 

 

Range/Resolution:

 20 Ω/0.01 Ω

Range/Resolution:

 200 Ω/0.1 Ω