Memmert IN30 Plus Incubator 32L

Set temperature range in °C     min. 5°C above ambient up to +80°C
Setting accuracy temperature     0.1°C
Temperature     2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Internals     1 stainless steel grid(s), electropolished
Works calibration certificate     incl. works calibration certificate for +37°C
Door     fully insulated stainless steel door with 2-point locking (compression door lock)
Door     inner glass door
ControlCOCKPIT     TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting     German, English, Spanish, French, Polish, Czech, Hungarian
Timer     Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT     the process time does not start until the set temperature is reached
Calibration     three freely selectable temperature values
adjustable parameters     temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Sterilisation     fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load
Interior     easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume     32 l
Dimensions W x H x D in mm     w(A) x h(B) x d(C): 400 x 320 x 250 mm
Max. number of internals     3
Max. loading of chamber     60 kg
Max. loading per internal     20 kg
Convection     natural convection
Fresh air admixture     adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually
Vent     vent connection with restrictor flap
Temperature control     mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Temperature control     overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY     additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system     for fault analysis
Alarm     visual and acoustic