Advanced Energy Innova 1403 多点采样仪 (20kPa, 2m/s)

Sampling System:
- Working Pump Suction: 20kPa
- Volume Flow Rate: 15ml/s
- Sample Transport Speed: 2m/s
- Minimum Pressure: 40 kPa (blocked airways)
Dosing System:
- Minimum Working Pump Pressure: 10 kPa Volume flow rate of supplementary air per dosing channel: 4 ml/s
- Volume Flow Rate: The volume flow rate is variable and is dependent on the selected tracer gas
- Max Delivery Time: 1 min
- Accuracy of Dosing Calculation: ±2%
Environmental:
- Operating Temperature: 5 ~ 40°C (41 ~ 104°F)
- Storage Temperature: -25 ~ 55°C (-13 ~ 131°F)
- Humidity: Max relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity @ 40°C
- Dimensions (W x H x D): 445 mm x 155 mm x 260 mm (17.5" x 6.10" x 10.2")
- Weight: 10 kg (22 lb)
Electrical:
- Power Requirement: 100 ~ 240 VAC 50 ~ 60 Hz
- Power Consumption: 0.9 A