Size Range: 0.3 - 25.0 µm
Flow Rate: 28.3 LPM (1.0 CFM) ±5%
Zero Count: < 7.07 counts / m3
Maximum concentration: > 42,924,662 / m3
Channels: Up to 8 configurable channels. Optional 2-channel pharmaceutical mode.
Counting Efficiency: 50% ± 20% for most sensitive channel. ISO 21501-4:2018 compliant.
Tubing Length: Maximum 8 meters
Compressed Gas: Air (CDA), argon (Ar), nitrogen (N), and carbon dioxide (CO2). Requires HPD III accessory.
Display and Printer: 8.4-inch color SVGA display, p-cap with finger swipe and gesture control. Optional built-in thermal printer with paper cutter. Supports cleanroom paper.
Exterior Surface: Polycarbonate (PC)
Data Storage: 10,000 complete data sets. Data cannot be altered and meets 21 CFR Part 11 compliance.
Reporting: Encrypted PDF, cleanroom monitoring and certification for ISO 14644-1, EU GMP, China GMP, and FS209E
Remote Operation: Real-time download to Facility Net, Pharmaceutical Net Pro, FacilityPro® SMART NG, or FacilityPro SCADA NG software, USB download, automatic clock adjustment with NTP server
Communications: Ethernet, USB data download, NFC tag for user login and sample recipe configuration
Input/Output: Four 4-20 mA inputs (<0.5% of Span), four 4-20 mA outputs (<1.0% Span), one 5V pulse output
Status/Alarm Output: 24 VDC, 1 A maximum
Operating Environment:
Temperature: 0 - 32 °C (32 - 89.6 °F)
Humidity: 5 - 95% RH non-condensing
External Dimensions: 24.6 x 25.7 x 25.6 cm (9.7 x 10.1 x 10.1 in)
Weight: 6.4 kg (14.2 lb) with two optional batteries; 5.4 kg (11.9 lb) without batteries
Power: 100 - 240 V, 50-60 Hz, 2.5 A. Up to two rechargeable Li-ion batteries.
Laser Classification: Class 1, complies with US 21 CFR 1040.10 and EN60825-1.
An internal: Class 4 laser is used per EN60825-1.
Calibration: Meets ISO 21501-4:2018 requirements. Optional ISO 17025 compliance available.