Rigel medical Multi-Flo Infusion Pump Analyzers (2 Channel, 0.100 ~ 1450ml/h)

Flow Measurement

Display Range: 0.010 to 1500ml/h

Max. Display Resolution: >10μl/h

Measured Range: 0.100 to 1450ml/h


Occlusion/Back Pressure Measurement

Pressure Measurement Range: -500 to 2500mmHg

Back Pressure Setting Range: -200 to 600mmHg

Accuracy:

-500 to 1000mmHg: ± 10mmHg

1000 to 2500mmHg: ± 1% of reading

Unit: mmHg, PSI, Bar, kPa

Resolution: 1mmHg


Flow Rate & Accuracy

701-1500ml/hr: 1 + x)% of reading where x = 0.0033% x (flow rate - 700ml/hr)

10-700ml/hr: 1% of reading

0.1-9.9ml/hr: 1% of reading + 0.005 ml/hr

Volume: 0.001 to 9999ml

Flow Update Rate: 1Hz


PCA / Bolus Measurements (Volume)

Display Range: 0.1 to 100ml

Measuring Range: 0.5 to 100ml

Accuracy: ± 1% of the reading

Max. Resolution: 10μl

Basal Flow Rate: 1ml to 30ml/h

Pressure: Max. 2500mmHg


General Specifications

Storage: 15 x 24h of testing at 1-second sampling rate

Dimensions: 300 x 204 x 220mm / 11.8 x 8 x 8.6” (w x d x h)

Weight: 6kg / 13lbs (2 channel)

Mains Supply: 90-264VAC, 50/60Hz, 60W

Mains Cable: 1ml to 30ml/h

Storage Environment: 0–50°C / 32–122°F

Operating Conditions: 15–40°C / 59–104°F

Environmental Protection: IP40

PC Communication: USB B

Keyboard Communication: USB A

Display: LCD color graphic display ¼” VGA

Details

Datasheet


Manual

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment