Huatec HBRV-187.5S 数字硬度计
制造商: HUATEC Model: HBRV-187.5S 来源: China - 联系
Technical specifications
1. The Power Source and the Voltage: AC220V±5%, 50-60 Hz
2.Time-delayed control: 0-60 seconds, adjustable
3.The Distance from the Indenter Center to the Instrument Body: 165mm
4.Overall Dimension (Length×Width×Height): 551×260×800 mm
5.The Net Weight of the Tester: 80kg (Approx)
Technical specifications
1. The Power Source and the Voltage: AC220V±5%, 50-60 Hz
2.Time-delayed control: 0-60 seconds, adjustable
3.The Distance from the Indenter Center to the Instrument Body: 165mm
4.Overall Dimension (Length×Width×Height): 551×260×800 mm
5.The Net Weight of the Tester: 80kg (Approx)
6.Specifications of Rockwell Hardness
Testing force(N) |
Initial Testing Force |
98.07(10kg) |
Tolerance ±2.0% |
|
Total testing force |
588.4(60kg) |
Tolerance ±1.0% |
||
980.7(100kg) |
||||
1471(150kg) |
||||
Indenter |
Diamond Cone Indenter |
|||
Ф1.5875mm Ball Indenter |
||||
Scales |
HRA |
HRB |
HRC |
HRD |
Max height of samples |
175mm |
7.The Tolerance of Rockwell Hardness Display Value
Hardness Scale |
HardnessRangeof The Standard Testing Blocks |
The Max. Tolerance of Display Value |
Repeatabilitya |
HRA |
(20~≤75)HRA |
±2HRA |
≤0.02(100-H)or 0.8 Rockwell Unitb |
(>75~≤88)HRA |
±1.5HRA |
||
HRB |
(20~≤45)HRB |
±4HRB |
≤0.04(130-H)or 1.2 Rockwell Unitb |
(>45~≤80)HRB |
±3HRB |
||
(>80~≤100)HRB |
±2HRB |
||
HRC |
(20~≤70)HRC |
±1.5HRC |
≤0.02(100-H)or 0.8 Rockwell Unitb |
a:H is average hardness value b:Confirm as higher value |
8.The Technical Specification of Brinell Hardness Tester
Testing force |
294.2N(30kg) |
Tolerance ±1.0% |
||
306.5N(31.25kg) |
||||
612.9N(62.5kg) |
||||
980.7N(100kg) |
||||
1839N(187.5kg) |
||||
Indenter |
φ2.5mm,φ5mmBall Indenter |
|||
Scales |
HBW1/30 |
HBW2.5/31.25 |
HBW2.5/62.5 |
|
HBW5/62.5 |
HBW10/100 |
HBW2.5/187.5 |
||
Eyepiece magnification |
15× |
|||
Objective |
2.5×(resolution 0.5μm),5×(resolution 0.25μm) |
|||
Max Height of Sample |
For 2.5×:Max Height is 95mm For 5×:Max height is 115mm |
9. The Tolerance and Repetition of Displayed Value for Brinell Hardness Tester
Hardness Block(HBW) |
Tolerance(%) |
Repeatability (%) |
≤125 |
±3 |
3 |
125<HBW≤125 |
±2.5 |
2.5 |
>225 |
±2 |
2 |
10. Technical Specifications of Vickers Hardness
Test force |
294.2N(30kg) |
Tolerance ±1.0% |
980.7N(100kg) |
||
Indenter |
Diamond Vickers Indenter |
|
Scale |
HV30 |
HV100 |
Eyepiece magnification |
15× |
|
Objective magnification |
5×(Resolution 0.25μm) |
|
Max. Height of Specimen |
115mm |
11. The Tolerance and Repetition of Displayed Value for Vickers Hardness Tester
Tolerance of Displayed Value |
Repetition of Displayed Value |
|||
Hardness scale |
Value of Hardness Block |
Tolerance of Displayed Value |
Value of Hardness Block |
Repetition of Displayed Value |
HV30 HV100 |
≤250HV |
±3% |
≤225HV |
6% |
300~1000HV |
±2% |
>225HV |
4% |
- 质量承诺
- 正品保修
- 送货到家
- 交易简单化